DropSenseฎ 96
Polychromatic ์l plate reader
Designed for reading DropPlates and microplates
UV VIS absorption spectrometry perfect fit for biomolecule
quantification
Full spectrum in seconds for multiprobe measurements
Large dynamic range (2 3500 ng/์l dsDNA) no need for
diluting concentrated samples
Easy to control Ready to use
Predefined methods for DNA/RNA and protein measurements
High reproducibility internal lamp compensation
Full developed software on remote PC

Key Applications
Determination of DNA/RNA purity and concentration in the range of 2 3500 ng/์l (dsDNA)
without dilution
Measuring the dye labelling efficiency of micro-array samples
Unique patent pending optical spectrometer design combining cylindrical lenses with integrated
optical elements, high resolution gratings and ultrasensitive photodiode array.
DropPlateฎ 16/96
0.5 2.5 μl sample quantities minimizing consumption
of your samples.
16 to 96 sample disposable technology your route
to higher throughput and automation
Minimized sample evaporation, allowing time and
temperature depending assays
Dynamic range: 2 3.500 ng/μl dsDNA
No sample dilution needed

Step 1
Fluid is dispensed into input well. Dispensing is done manually or via a liquid
handling robot. The chip is not in the DropSense system during dispensing.
Step 2
Fluid is drawn into capillary channel due to capillary action (hydrophilized COC
material). The chip is now ready to be loaded into the system.
A first optical measurement will be performed to verify the absorption of the chip
material. No sample is yet in the optical pathway.
Step 3
First chamber is filled by applying a small vacuum pressure at the evacuation hole
above the spill chamber. This pressure is supplied by an external source, integrated
in the DropSense system. Another optical measurement is performed
Step 4
Second chamber is also filled (first chamber remains also filled) by applying a
higher vacuum pressure. Another optical measurement is performed

User Interface
Easy to operate
Predefined protocols
Sample identification
Multiple user levels
Interface to robotic systems
LIMS ready
Database/Lims compatible
Specifications:
Sample size: 0,5 2,5 microliter
Sample number: 1 16 96
Path lengths: 0.2 mm and 1mm
Light source: Xenon flash lamp
Optical system: 4-channel polychromator
+ reference channel
Wavelength range: 220 750 nm
Wavelength resolution: 5 nm
Wavelength accuracy: 0,5 nm
Wavelength reproducibility: 0.1 nm
Absorbance range: OD 0.05 90
(10mm equivalent absorbance)
Detection limit: 2 ng/microliter (dsDNA)
Maximum concentration: 3500 ng/microliter (dsDNA)
Measurement cycle time: < 4 minutes for 96 samples
Dimensions (footprint): 34 x 45 x 35 cm
Operating voltage: 24 VDC
Connection: USB
CE approval: All units
|